Off-center load cell TA11


Download PDF

Off-center load cell for weighing systems, automatic dosage machines, medical, food, textile, and automation in general

Off-center load cell

Upon request with certification ATEX or IECEx.


Zoom Click to enlarge image

Off-center load cell TA11 Off-center load cell TA11 Schema
ATEX IECEX

Technical Specifications

MaterialAluminium
Nominal load0.6, 1, 3, 5, 6, 10 kg
Limit load150 % F.S.
Breaking load300 % F.S.
Supply voltageVcc 15 Max
Output1-2 mV/V
Output tollerance0,1 % F.S.
Zero tollerance2 % F.S.
Linearity0,02 % F.S.
Hysteresis0,02 % F.S.
Reapeatability0,02 % F.S.
Input resistance400 – 410 ohm
Output resistance350-353 ohm
Insulation resistance>= 2000 Mohm
Creep (30 Minutes)0,03 % F.S.
Thermal compensation-10 / +40°C (14 / +104°F)
Operating temperature-20 / +70°C (-4° / +158°F)
Temperature deviation zero0,003 % F.S. / °C
Temperature deviation full scale0,002 % F.S. / °C
Protection classIP 65
Standard cable length0,4 - 2,0 m – 4 x 0,15 (shield not connected to the load cell body)
For the complete and latest technical specifications, please refer to the technical data sheet which can be downloaded here:
Download PDF

Connections

TypeSHIELDED CABLE PVC 4C R3 4X0,15
Length0,4 - 2 m (body lead cell not connected shield)
Red+ IN
Black- IN
Green+ OUT
White- OUT
Connections  Off-center load cell TA11

MOUNTING OF THE LOAD CELL

General precautions to be followed in the installation and assembly of load cells:

  • Follow the direction of application of force on the load cell
  • Do not exceed the limits of the temperature as indicated in the data sheet
  • Do not remove labels from the load cell
  • Adequately protect the cables with sleeves or channels
  • Carry out on each load cell a ground connection with suitable cable, shunt on mounting accessories
  • Do not carry out any welding on the mechanical structure after mounting the load cell
  • Avoid or reduce to a minimum level any mechanical constraints between the rigid structure and the weighing structure to avoid error in weighing and return to zero
  • In case of vibration or dynamic loads product the load cells with puffers / shock absorbers
  • Protect load cell from any dust accumulation with any appropriate means